Audited ·Last updated 29 Jul 2026·6 citations·Tier 1·0 uses

DNA Concentration Calculator — A260 to ng/µL

Convert A260 absorbance to ng/µL for dsDNA, ssDNA or RNA, with total yield, 260/280 and 260/230 purity ratios and the pH caveat that changes them.

DNA Concentration Calculator (A260)

What did you measure?
Absorbance at 260 nm, in absorbance units, as read. Keep the reading between roughly 0.1 and 1.0 AU — outside that window most photometers are no longer linear, and you should dilute or concentrate rather than trust the number.
AU
Absorbance at 280 nm — the aromatic-amino-acid peak, used for the protein-carryover ratio. Leave it at 0 if you did not measure it; the 260/280 ratio is then reported as not evaluated instead of dividing by zero.
AU
Absorbance at 230 nm — sensitive to guanidine salts, EDTA, phenol and carbohydrate. Leave it at 0 if you did not measure it.
AU
Fold dilution of the material you actually put in the cuvette. 1 = read neat. 10 µL of sample plus 90 µL of buffer is 10. The result is reported for the ORIGINAL undiluted sample.
×
Optical path length of the measurement. A standard cuvette is 1 cm. Microvolume pedestal instruments physically use 1 mm or 0.2 mm but report a 10 mm-equivalent absorbance, so leave this at 1 unless your instrument tells you otherwise.
cm
Used only when the analyte above is set to Custom. Enter the value from your own extinction table, or a sequence-specific factor for a short oligonucleotide. Published tables disagree for single-stranded DNA — 33, 37 and 40 all appear in print.
(µg/mL)/AU
Volume of the whole undiluted sample — your elution volume, typically. Used only for the total-yield figure.
µL
How much DNA or RNA the next reaction needs. The calculator returns the pipetting volume that delivers it.
ng
Concentration
206
Concentration of the original, undiluted sample: (A260 ÷ path length) × conversion factor × dilution factor. Numerically identical to µg/mL. Report it to about 3 significant figures — that is all a photometer resolves.
Total yield
10.3 µg
A260/A280
1.8643
A260/A230
2.0808
Volume for target mass
2.4272 µL
Conversion factor used
50 (µg/mL)/AU
Reading the ratios
260/280 = 1.86 — inside the 1.70–2.00 band conventionally quoted for pure DNA. 260/230 = 2.08 — inside the 2.00–2.20 band. Both ratios are assay-dependent: an acidic diluent under-reports 260/280 by 0.2–0.3 and an alkaline one over-reports it by the same amount, which is wider than these bands. Treat a ratio outside the band as a prompt to re-read in 10 mM Tris·HCl pH 8.0–8.5, not as a purity verdict.

Background.

This calculator turns a UV absorbance reading at 260 nm into a nucleic acid concentration in ng/µL, a total yield in µg, the two purity ratios everyone quotes, and the pipetting volume that delivers whatever mass your next reaction needs. It is the arithmetic behind every spectrophotometer readout — written out so you can see which constant was applied and change it if your lab uses a different one.

The physics is Beer–Lambert: absorbance is proportional to concentration and to the distance light travels through the sample, A = ε·c·l. Rearranged for concentration, c = A ÷ (ε·l). Rather than carry a molar extinction coefficient, nucleic acid work uses a mass-based shortcut — a conversion factor in micrograms per millilitre per absorbance unit per centimetre. For double-stranded DNA that factor is 50, for single-stranded DNA 33, and for RNA 40. Those three numbers are the ones tabulated by Desjardins and Conklin in the Journal of Visualized Experiments, and they are what a NanoDrop-class instrument applies when you press the button. Double-stranded DNA has the lowest factor per microgram because base stacking inside an intact duplex suppresses absorbance — the hypochromic effect — so the same mass of DNA absorbs less when it is paired than when it is melted.

The conversion factor is the whole reason a page like this exists. It assumes a long, compositionally random polymer. It is a poor approximation for a 20-mer primer, for a homopolymer stretch, and for anything with unusual base composition, because the absorbance of a short oligonucleotide depends on which bases sit next to which. If you are quantifying an oligo, use a sequence-specific extinction coefficient from your supplier and enter it in the Custom field — do not use 33 and hope. Published tables also disagree with each other for single-stranded DNA: 33, 37 and 40 all appear in print, from different lineages of the same underlying measurement. This calculator uses 33 because it is the value the instrument that generated your reading almost certainly applied, and it exposes the field so you are not trapped in that choice.

The two ratios need more care than they usually get. A260/A280 is conventionally quoted as about 1.8 for pure DNA and about 2.0 for pure RNA, and A260/A230 as somewhere between 2.0 and 2.2. Those are conventions reported by instrument manufacturers, not specifications published by a standards body, and the same documents that publish them also report that moving the diluent from acidic to alkaline shifts A260/A280 by 0.2 to 0.3 in either direction. That artefact is wider than the band. Wilfinger, Mackey and Chomczynski measured the effect directly in 1997 and found RNA ratios moving from roughly 1.5 to 2.0 simply by taking the water from pH 5.4 to pH 7.5–8.5. So a ratio outside the band is a prompt to re-read the sample in 10 mM Tris·HCl at pH 8.0–8.5 before you conclude anything about purity. The calculator prints that caveat with every result rather than hiding it in an FAQ.

One more scope limit belongs up here rather than below the fold: absorbance measures how much material absorbs at 260 nm, and nothing else. It cannot tell an intact 20 kb genomic fragment from the same mass of sheared 200 bp debris, and it happily counts free nucleotides, degraded RNA and carry-over nucleic acid of the wrong type. This is why NIST assigns certified values to its human genomic DNA standard by PCR-based copy-number and mass-concentration methods rather than by absorbance, and why a fluorescent dye assay is the right tool when you need to know how much of a specific analyte is present in a mixed sample. Use A260 for a fast, non-destructive, reagent-free estimate of total nucleic acid. Use a dye-based assay when the number has to be right.

Everything below the widget — the derivation, the worked example, the ratio interpretation table and the FAQs — is generated from the same worked case the calculator computes, so the numbers in the prose and the numbers on screen cannot drift apart.

What is dna concentration calculator (a260)?

A260 quantification is the measurement of a nucleic acid solution's absorbance at 260 nanometres, the wavelength at which the purine and pyrimidine bases absorb most strongly, and its conversion into a mass concentration using an empirical factor. The relationship is the Beer–Lambert law: A = ε·c·l, where A is absorbance (dimensionless), ε is the extinction coefficient, c is concentration and l is the optical path length. Because ε for a nucleic acid depends on base composition and on whether the strands are paired, wet-lab practice replaces the molar coefficient with a mass-based conversion factor expressed in micrograms per millilitre per absorbance unit at a 1 cm path. Those factors — 50 for double-stranded DNA, 33 for single-stranded DNA, 40 for RNA — are averages over random sequence. An absorbance of 1.000 AU at 260 nm through a 1 cm path therefore means 50 ng/µL of double-stranded DNA, 33 ng/µL of single-stranded DNA, or 40 ng/µL of RNA. Two secondary readings are taken at the same time. A280 sits near the absorbance maximum of the aromatic amino acids tryptophan and tyrosine, so the A260/A280 ratio falls when protein is carried over. A230 sits where guanidine salts, EDTA, phenol and carbohydrates absorb, so the A260/A230 ratio falls when extraction reagents are carried over. Neither ratio is a concentration and neither is a specification: they are diagnostics whose value shifts measurably with the pH and ionic strength of the diluent, independently of how pure the sample actually is. Microvolume pedestal spectrophotometers complicate one detail worth knowing. They hold the sample in a liquid column typically 1 mm or 0.2 mm deep rather than in a 1 cm cuvette, then scale the raw absorbance to a 10 mm-equivalent value before reporting it. That is why you normally leave the path length in this calculator at 1 cm even when the physical path was a tenth of that — the instrument has already done the conversion.

How to use this calculator.

  1. Choose the analyte. Double-stranded DNA (genomic preps, plasmid minipreps, PCR products, restriction fragments) uses 50. Single-stranded DNA uses 33. RNA — total RNA, mRNA, in vitro transcripts — uses 40. Pick Custom for a short oligonucleotide and enter the sequence-specific factor from your supplier's spec sheet.
  2. Enter A260 exactly as the instrument reported it. If the reading is above about 1.0 AU or below about 0.1 AU, dilute or concentrate and read again rather than trusting an extrapolation outside the linear range.
  3. Enter A280 and A230 if you measured them. Leave either at 0 if you did not — the calculator reports that ratio as 'not evaluated' rather than dividing by zero and returning nonsense.
  4. Set the dilution factor to the fold dilution of the material that went into the cuvette. Reading neat is 1. Ten microlitres of sample plus ninety of buffer is 10. The concentration reported is always that of the original undiluted sample.
  5. Leave the path length at 1 cm for a standard cuvette and for microvolume pedestal instruments, which already normalise to a 10 mm equivalent. Change it only if your instrument reports raw absorbance at a physical short path.
  6. Enter your elution volume as the sample volume to get the total yield of the whole prep in micrograms.
  7. Enter the mass your next reaction needs — a ligation, a sequencing submission, a library prep — and the calculator returns the microlitres of this sample that deliver it.
  8. Read the ratios as flags, not grades. If either is outside its band, re-read the sample in 10 mM Tris·HCl at pH 8.0–8.5 before you re-purify anything: the pH artefact alone is bigger than the band.

The formula.

c = (A₂₆₀ ⁄ l) × F × D

Four equations run this page, and all of them are one rearrangement away from Beer–Lambert.

c[µg/mL] = (A260 / l) × F × D concentration of the original sample Y[µg] = c[ng/µL] × V[µL] / 1000 total yield of the whole prep R280 = A260 / A280 protein-carryover ratio R230 = A260 / A230 reagent-carryover ratio Vneed = m_target[ng] / c[ng/µL] pipetting volume for a target mass

The first line is A = ε·c·l solved for c, with the molar extinction coefficient replaced by a mass-based factor F. Dividing by the path length converts a reading taken through any optical geometry to the 1 cm reference, and multiplying by the dilution factor D undoes whatever dilution you made before reading, so the answer describes the tube on your bench rather than the cuvette.

The unit identity in the first line is exact, not an approximation, and it is worth knowing because it removes a whole class of conversion errors: 1 µg/mL = 10⁻⁶ g ÷ 10⁻³ L = 10⁻³ g/L, and 1 ng/µL = 10⁻⁹ g ÷ 10⁻⁶ L = 10⁻³ g/L. They are the same quantity written two ways. A NanoDrop reporting 206 ng/µL and a cuvette calculation giving 206 µg/mL agree exactly; neither needs converting.

Rounding stage: FINAL ONLY. Every intermediate — the division by path length, the two multiplications, both ratios, the yield and the pipetting volume — is carried at full precision in Decimal arithmetic, and rounding to ten decimal places happens once, at the moment the result is returned. Nothing is rounded and then reused. The scientific precision of the answer is set by the photometer, not by the arithmetic: a benchtop instrument resolves about three significant figures, so 206 ng/µL is the honest way to report the worked example below, and quoting 206.0000000000 would be a lie about the measurement.

Invalid-domain behaviour is deliberate and worth stating. A negative absorbance is rejected outright, because it is a blanking artefact rather than a datum — the fix is to re-blank against your elution buffer, not to carry a negative concentration forward. A path length or dilution factor of zero is rejected because both are divisions by zero with no physical reading. An A280 or A230 of zero is NOT rejected: 'I only measured A260' is a legitimate workflow, so the corresponding ratio returns zero and the text output states plainly that it was not evaluated. Likewise an A260 of zero gives a concentration of zero, and the pipetting volume that would deliver a target mass from a zero-concentration sample is unbounded, so it returns zero with an explicit note rather than an infinity.

One approximation runs underneath all of it. The factors 50, 33 and 40 are averages over random sequence, valid for long polymers. They are not valid for short oligonucleotides, homopolymer runs, or sequences with strongly skewed base composition, where the absorbance depends on nearest-neighbour stacking and a sequence-specific coefficient is required. That is what the Custom field is for.

A worked example.

Example

A genomic DNA miniprep eluted in 50 µL. You dilute 10 µL of it into 90 µL of elution buffer — a 10-fold dilution — and read the dilution in a 1 cm cuvette: A260 = 0.412, A280 = 0.221, A230 = 0.198. Double-stranded DNA, so the conversion factor is 50 (µg/mL)/AU. Concentration first: (0.412 ÷ 1) × 50 × 10 = 206 µg/mL, which is 206 ng/µL of the original undiluted eluate. Note the dilution factor pulling the number up, not down — the cuvette held a tenth of the stock, so the stock is ten times more concentrated than what you read. Total yield: 206 ng/µL × 50 µL = 10 300 ng, which is 10.3 µg from the whole prep. Ratios next. A260/A280 = 0.412 ÷ 0.221 = 1.86, inside the 1.70–2.00 band conventionally quoted for pure DNA. A260/A230 = 0.412 ÷ 0.198 = 2.08, inside the 2.00–2.20 band. Both readings look clean — but the pH caveat still applies, because a diluent one or two pH units off could have moved that 1.86 anywhere between roughly 1.6 and 2.1 without a single molecule of protein being present. Finally, suppose the library prep downstream needs 500 ng of input. 500 ng ÷ 206 ng/µL = 2.43 µL. That is an awkward volume to pipette accurately with a P2, so in practice you would dilute an aliquot of the eluate tenfold to 20.6 ng/µL and pipette 24.3 µL instead — same mass, four times the pipetting precision. Report the concentration as 206 ng/µL, three significant figures. The calculator will show more decimal places because the arithmetic is exact; the measurement is not.

a2300.198
a2800.221
a2600.412
dilution Factor10
nucleic Acid TypedsDNA
path Length1
target Mass500
sample Volume50
custom Factor33

Frequently asked questions.

Why is the conversion factor 50 for double-stranded DNA but only 33 for single-stranded DNA?
Because paired bases absorb less light than unpaired ones. In an intact double helix the bases are stacked and electronically coupled, which suppresses absorbance at 260 nm by roughly 30–40 percent relative to the same bases free in solution. This is the hypochromic effect, and it is the basis of every melting-curve experiment: as a duplex denatures, absorbance rises. The practical consequence is that a microgram of double-stranded DNA gives a smaller A260 than a microgram of single-stranded DNA, so it takes more micrograms of dsDNA to reach an absorbance of 1.000 — 50 rather than 33. RNA sits at 40 because it is single-stranded but heavily structured, with substantial intramolecular base pairing that recovers some of the hypochromicity.
My 260/280 ratio is 1.6. Is my DNA contaminated with protein?
Possibly, but you cannot tell from that number alone, and the first thing to check is not your prep but your diluent. Wilfinger, Mackey and Chomczynski showed in 1997 that simply moving the water used for the measurement from pH 5.4 to pH 7.5–8.5 raised RNA 260/280 ratios from about 1.5 to about 2.0 — a swing of 0.5 with no change in sample purity whatsoever. Instrument manufacturers quote a smaller but still substantial effect: acidic solutions under-report the ratio by 0.2–0.3, alkaline solutions over-report it by the same margin. So re-read the sample in 10 mM Tris·HCl at pH 8.0–8.5 before concluding anything. If the ratio stays low after that, then look at protein carryover, phenol (which absorbs strongly near 270 nm and drags the ratio down hard), or guanidine. The 260/230 ratio helps discriminate: protein depresses 260/280 preferentially, while phenol and chaotropic salts hit 260/230 much harder.
Can I trust A260 quantification for a PCR primer or a short oligonucleotide?
Not with the generic 33 factor. The mass conversion factors are averages over long, compositionally random polymers. For a short oligo, absorbance depends strongly on which bases sit next to which — nearest-neighbour stacking interactions change the extinction coefficient by well over ten percent between sequences of identical length and identical base composition. Every reputable oligo supplier ships a sequence-specific extinction coefficient with the synthesis report, usually in L·mol⁻¹·cm⁻¹, and that is the number to use. If you have it as a mass factor in (µg/mL)/AU, put it in the Custom field on this page. If you only have the molar coefficient, convert the absorbance to molarity with c = A/(ε·l) and then to mass with the oligo's molecular weight — our primer GC content calculator computes that molecular weight from the sequence.
Does the dilution factor make the reported concentration higher or lower?
Higher. The dilution factor describes what you did to the sample before reading it, so the calculator has to undo it. If you diluted 10 µL of eluate into 90 µL of buffer, the cuvette held material at one tenth the stock concentration, and the stock must therefore be ten times whatever the cuvette reading implies. In the worked example above, the cuvette reading of A260 = 0.412 corresponds to 20.6 ng/µL in the cuvette and 206 ng/µL in the tube. The commonest error here is entering the dilution as 0.1 instead of 10 — a fold dilution is always a number greater than or equal to 1.
Why does my microvolume instrument use a 1 mm path but the calculator says 1 cm?
Because the instrument has already converted. Pedestal spectrophotometers hold the sample in a short liquid column — commonly 1 mm, dropping to 0.2 mm automatically for concentrated samples — and then scale the measured absorbance up to what a 10 mm cuvette would have read, so that the number on the screen is directly comparable with cuvette literature and with the standard conversion factors. Leave the path length at 1 cm unless your instrument documentation explicitly says it reports raw absorbance at the physical path length. If it does, enter the physical path in centimetres: 1 mm is 0.1 cm, and the calculator will scale accordingly.
What A260 reading is actually reliable?
Roughly 0.1 to 1.0 absorbance units on a cuvette instrument. Below about 0.1 AU the signal is close to the noise floor and to the drift of the blank, so small baseline errors become large percentage errors in concentration. Above about 1.0 AU most photometers leave their linear range — stray light and detector saturation both compress the reading, so the true absorbance is higher than reported and your concentration comes out low. Microvolume instruments extend the top end considerably by switching to a shorter physical path, which is the whole point of the design, but they are still worse than a cuvette at the bottom end because the sample column is small and evaporation matters. If your reading is outside the window, dilute or concentrate and read again; do not extrapolate.
Should I use A260 or a fluorescent dye assay?
Use A260 when you want a fast, free, non-destructive estimate of total nucleic acid and you know roughly what is in the tube. Use a dye assay — the intercalating and groove-binding chemistries sold for exactly this — when the number has to be specific. Absorbance at 260 nm cannot distinguish your analyte from anything else that absorbs there: free nucleotides left over from a PCR, degraded RNA in a DNA prep, carrier RNA deliberately added during extraction, and single-stranded oligos all count toward the reading. A dsDNA-specific dye counts only intact double-stranded DNA. This is not a subtle difference: an RNA-contaminated genomic prep can read 30–50 percent high by A260 and correctly by dye. It is also why NIST certifies its human genomic DNA reference material by PCR-based copy-number and mass-concentration methods rather than by absorbance.
What does a 260/230 ratio below 2.0 mean?
It means something in the sample absorbs near 230 nm, and the usual suspects are all extraction reagents rather than biological contaminants. Guanidine thiocyanate and guanidine hydrochloride from silica-column chemistries are the most common cause and the easiest to fix — an extra wash with the ethanol buffer, and a proper dry spin before elution, usually recovers the ratio. EDTA at high concentration does it too, which is why eluting in 10 mM Tris rather than TE matters if you care about this ratio. Phenol depresses both 260/230 and 260/280. Carbohydrate carryover from plant and some bacterial preps is a stubborn one. A low 260/230 with a normal 260/280 usually points at salts and solvents; both ratios low usually points at phenol or protein. As with 260/280, check that A230 itself is not simply tiny and noisy before reading anything into a ratio built on it.
Can this calculator tell me if my DNA is degraded?
No, and no absorbance measurement can. Beer–Lambert relates absorbance to the amount of absorbing material, not to how that material is arranged. A 20 kilobase genomic fragment and the same mass of it sheared into 200 base-pair pieces give identical A260 readings, identical concentrations, and identical purity ratios. If fragment length matters — and for long-read sequencing, Southern blotting or large-insert cloning it matters a great deal — you need a method that separates by size: an agarose gel, a capillary or microfluidic sizing instrument, or a pulsed-field gel for very high molecular weight material. Treat the concentration from this page as 'how much', and use an orthogonal method for 'in what state'.
Which conversion factor should I use for single-stranded DNA — 33, 37 or 40?
All three appear in print, from different lineages of the same underlying measurement, and this calculator defaults to 33. That value is tabulated by Desjardins and Conklin and is what microvolume instruments of the NanoDrop class apply, so if your reading came off such an instrument, 33 keeps your hand calculation consistent with the machine's. The figure 37 comes from tables that quote an extinction coefficient of 0.027 (µg/mL)⁻¹cm⁻¹ for single-stranded DNA, whose reciprocal is 37.0. The figure 40 appears in older manual-lineage tables that use one value for both single-stranded DNA and RNA. The spread between 33 and 40 is about 20 percent, which is larger than most people assume their quantification error to be — so if the exact number matters, state which factor you used in your methods section, and use the Custom field here to match whatever your lab has standardised on.

References& sources.

  1. [1]Desjardins, P. & Conklin, D. (2010). 'NanoDrop Microvolume Quantitation of Nucleic Acids.' Journal of Visualized Experiments 45:e2565, doi:10.3791/2565. PRIMARY SOURCE, peer-reviewed, open access. Table 1 gives the conversion constants used here — DNA-50 = 50, DNA-33 = 33, RNA-40 = 40 (µg/mL per absorbance unit) — and states that pure nucleic acids typically yield a 260/280 ratio of ~1.8 for DNA and ~2.0 for RNA, with acidic solutions under-representing the ratio by 0.2–0.3 and basic solutions over-representing it by 0.2–0.3. Not independent of the NanoDrop instrument lineage; verified against a competing manufacturer below. Retrieved 2026-07-29.
  2. [2]Wilfinger, W. W., Mackey, K. & Chomczynski, P. (1997). 'Effect of pH and ionic strength on the spectrophotometric assessment of nucleic acid purity.' BioTechniques 22(3):474–481, doi:10.2144/97223st01, PMID 9067025. INDEPENDENT, peer-reviewed. Measured RNA A260/A280 rising from approximately 1.5 to 2.0 on moving the diluent from pH ~5.4 to pH 7.5–8.5, and recommends measuring in 1–3 mM Na2HPO4 at pH 8.0–8.5 for reproducibility. Abstract verified 2026-07-29; full text is paywalled at Taylor & Francis.
  3. [3]DeNovix Technical Note TN 130, 'Purity Ratios Explained', revision date 2026-07-20. INDEPENDENT SECOND AUTHORITY — an instrument manufacturer separate from the NanoDrop lineage. Consulted specifically to check the primary source, and it agrees on every figure: 260/280 of ~1.8 for DNA and ~2.0 for RNA; 'generally acceptable 260/230 ratios are in the range of 2.0 – 2.2'; and the same ±0.2–0.3 pH artefact, which it independently attributes to Wilfinger et al. 1997. Retrieved and verified 2026-07-29.
  4. [4]Thermo Scientific NanoDrop Technical Bulletin T042, '260/280 and 260/230 Ratios' — manufacturer technical bulletin, public mirror hosted by the University of Vermont Genetics Network. Source of the 2.0–2.2 expectation for 260/230 as a secondary purity measure. PDF retrieved 2026-07-29; its text layer could not be machine-extracted, so the figures it carries are the ones independently verified in DeNovix TN 130 above. Corroborating reference, not the load-bearing one.
  5. [5]National Institute of Standards and Technology, Standard Reference Material 2372a, 'Human DNA Quantitation Standard' — certificate of analysis. STANDARDS BODY. Cited for scope rather than for a constant: NIST assigns certified values of DNA mass concentration and copy number to this material using PCR-based methods, not UV absorbance, which is the clearest available statement that A260 is an estimate of total absorbing material rather than a traceable measurement of a specific analyte. Retrieved 2026-07-29.
  6. [6]International Union of Pure and Applied Chemistry, Compendium of Chemical Terminology (the 'Gold Book'), entries for 'absorbance' and the Beer–Lambert law. DEFINITIONAL AUTHORITY for A = ε·c·l and for absorbance being a dimensionless quantity referred to a stated path length. Automated retrieval of the Gold Book was blocked on 2026-07-29; the relation is standard and is restated identically in both manufacturer documents cited above.

In this category

Embed

Quanta Pro

Paid features are coming later.

  • All 682 calculators remain free
  • No billing is enabled
Coming soon